Full Description
HPLC UV100 Detector
Basic Technical Criteria
(1) Wavelength range: 190-600nm
(2) Wavelength indicated value error: from -2nm to +2nm
(3) Wavelength repetitiveness: better than 2nm
(4) Spectral bandwidth: 8nm
(5) Baseline noise: no more than 5 × 10-5AU(C18 chromatographic column, methanol1mL/min,254nm)
(6) Baseline drift: no more than 1 × 10-3AU/h (C18 chromatographic column, methanol1mL/min,254nm, after the instrument runs stably)
(7) Minimum detectable concentration: no more than 2 × 10-8 g/ml(naphthalene/methanol solution)
(8) Linear range: more than 103
(9) Time constant of filtration (second):
0.1 0.2 0.5 1 2
(10) Physical dimension: 450 mm ×300 mm ×160 mm (length×width×height)
(11) Weight: 12kg
(12) Power supply: AC(220±22)V; frequency (50±0.5)Hz; or
AC (110±11) V and (60±0.5) Hz (in need of switching start/stop)